|
Arizona
Data Qualifiers
Revision 1.0
05/13/2002
Developed by the Sub-committee of the Arizona Environmental Laboratory
Advisory Committee.
This
is a revised list with additional qualifiers added to the original list dated
12/11/2000.
The new items are in bold.
Microbiology:
A1 =
Too numerous to count.
A2 =
Sample incubation period exceeded method requirement.
A3 =
Sample incubation period was shorter than method requirement.
A4 =
Target organism detected in associated method blank.
A5 =
Incubator/water bath temperature was outside method requirements.
A6 =
Target organism not detected in associated positive control.
A7 =
Micro sample received without adequate headspace.
Method
blank:
B1 =
Target analyte detected in method blank at or above the method reporting
limit.
B2 =
Non-target analyte detected in method blank and sample, producing
interference.
B3 =
Target analyte detected in calibration blank at or above the method reporting
limit.
B4 =
Target analyte detected in blank at/above method acceptance criteria.
B5 =
Target analyte detected in method blank at or above the method
reporting limit, but below trigger level or MCL.
B6 =
Target analyte detected in calibration blank at or above the method reporting
limit, but below trigger level or MCL.
B7 =
Target analyte detected in method blank at or above method reporting limit.
Concentration found in the sample was 10 times above the concentration found
in the method blank.
Confirmation:
C1 =
Confirmatory analysis not performed as required by the method.
C2 =
Confirmatory analysis not performed. Confirmation of analyte presence
established by site historical data.
C3 =
Qualitative confirmation performed. See case narrative.
C4 =
Confirmatory analysis was past holding time.
C5 =
Confirmatory analysis was past holding time. Original result not confirmed.
Dilution:
D1 =
Sample required dilution due to matrix interference. See case narrative.
D2 =
Sample required dilution due to high concentration of target analyte.
D3 =
Sample dilution required due to insufficient sample.
D4 =
Minimum reporting level (MRL) adjusted to reflect sample amount received and
analyzed.
Estimated
concentration:
E1 =
Concentration estimated. Analyte exceeded calibration range. Reanalysis not
possible due to insufficient sample.
E2 =
Concentration estimated. Analyte exceeded calibration range. Reanalysis not
performed due to sample matrix.
E3 =
Concentration estimated. Analyte exceeded calibration range. Reanalysis not
performed due to holding time requirements.
E4 =
Concentration estimated. Analyte was detected below laboratory minimum
reporting level (MRL).
E5 =
Concentration estimated. Analyte was detected below laboratory minimum
reporting level (MRL), but not confirmed by alternate analysis.
E6 =
Concentration estimated. Internal standard recoveries did not meet method
acceptance criteria.
E7 =
Concentration estimated. Internal standard recoveries did not meet laboratory
acceptance criteria.
Hold
time:
H1 =
Sample analysis performed past holding time. See case narrative.
H2 =
Initial analysis within holding time. Reanalysis for the required dilution was
past holding time.
H3 =
Sample was received and analyzed past holding time.
H4 =
Sample was extracted past required extraction holding time, but analyzed
within analysis holding time. See case narrative.
BOD:
K1 =
The sample dilutions set-up for the BOD analysis did not meet the oxygen
depletion criteria of at least 2 mg/L. Any reported result is an estimated
value.
K2 =
The sample dilutions set up for the BOD analysis did not meet the criteria of
a residual dissolved oxygen of at least 1 mg/L. Any reported result is an
estimated value.
K3 =
The seed depletion was outside the method acceptance limits.
K4 =
The seed depletion was outside the method and laboratory acceptance limits.
The reported result is an estimated value.
K5 =
The dilution water D.O. depletion was > 0.2 mg/L.
K6 =
Glucose/glutamic acid BOD was below method acceptance criteria.
K7 = A
discrepancy between the BOD and COD results has been verified by reanalysis of
the sample for COD.
K8=
Glucose/glutamic acid BOD was above method acceptance levels.
Laboratory
fortified blank/blank spike:
L1 =
The associated blank spike recovery was above laboratory acceptance limits.
See case narrative.
L2 =
The associated blank spike recovery was below laboratory acceptance limits.
See case narrative.
L3 =
The associated blank spike recovery was above method acceptance limits. See
case narrative.
L4 =
The associated blank spike recovery was below method acceptance limits. See
case narrative.
Note:
The L1, L2, L3 & L4 footnotes need to be added to all corresponding
analytes for a sample.
Matrix
spike:
M1 =
Matrix spike recovery was high, the method control sample recovery was
acceptable.
M2 =
Matrix spike recovery was low, the method control sample recovery was
acceptable.
M3 =
The accuracy of the spike recovery value is reduced since the analyte
concentration in the sample is disproportionate to spike level. The method
control sample recovery was acceptable.
M4 =
The analysis of the spiked sample required a dilution such that the spike
concentration was diluted below the reporting limit. The method control sample
recovery was acceptable.
M5 =
Analyte concentration was determined by the method of standard addition (MSA).
M6=
Matrix spike recovery was high. Data reported per ADEQ policy 0154.000.
M7=
Matrix spike recovery was low. Data reported per ADEQ policy 0154.000.
General:
N1 =
See case narrative.
N2 =
See corrective action report.
Sample
quality:
Q1 =
Sample integrity was not maintained. See case narrative.
Q2 =
Sample received with head space.
Q3 =
Sample received with improper chemical preservation.
Q4 =
Sample received and analyzed without chemical preservation.
Q5 =
Sample received with inadequate chemical preservation, but preserved by the
laboratory.
Q6 =
Sample was received above recommended temperature.
Q7 =
Sample inadequately dechlorinated.
Q8 =
Insufficient sample received to meet method QC requirements. QC requirements
satisfy ADEQ policies 0154 and 0155.
Q9 =
Insufficient sample received to meet method QC requirements.
Q10 =
Sample received in inappropriate sample container.
Q11 =
Sample is heterogeneous. Sample homogeneity could not be readily achieved
using routine laboratory practices.
Duplicates:
R1 =
RPD exceeded the method control limit. See case narrative.
R2 =
RPD exceeded the laboratory control limit. See case narrative.
R3 =
Sample RPD between the primary and confirmatory analysis exceeded 40%. Per EPA
Method 8000B, the higher value was reported.
R4 =
MS/MSD RPD exceeded the method control limit. Recovery met acceptance
criteria.
R5 =
MS/MSD RPD exceeded the laboratory control limit. Recovery met acceptance
criteria.
R6 =
LFB/LFBD RPD exceeded the method control limit. Recovery met acceptance
criteria.
R7 =
LFB/LFBD RPD exceeded the laboratory control limit. Recovery met acceptance
criteria.
R8 =
Sample RPD exceeded the method control limit.
R9 =
Sample RPD exceeded the laboratory control limit.
Surrogate:
S1 =
Surrogate recovery was above laboratory acceptance limits, but within method
acceptance limits.
S2 =
Surrogate recovery was above laboratory and method acceptance limits.
S3 =
Surrogate recovery was above laboratory acceptance limits, but within method
acceptance limits. No target analytes were detected in the sample.
S4 =
Surrogate recovery was above laboratory and method acceptance limits. No
target analytes were detected in the sample.
S5 =
Surrogate recovery was below laboratory acceptance limits, but within method
acceptance limits.
S6 =
Surrogate recovery was below laboratory and method acceptance limits.
Reextraction and/or reanalysis confirms low recovery caused by matrix effect.
S7 =
Surrogate recovery was below laboratory and method acceptance limits. Unable
to confirm matrix effect.
S8 =
The analysis of the sample required a dilution such that the surrogate
concentration was diluted below the method acceptance criteria. The method
control sample recovery was acceptable.
S9 =
The analysis of the sample required a dilution such that the surrogate
concentration was diluted below the laboratory acceptance criteria. The method
control sample recovery was acceptable.
S10 =
Surrogate recovery was above laboratory and method acceptance limits. See Case
narrative.
S11=
Surrogate recovery was high. Data reported per ADEQ policy 0154.000.
S12=
Surrogate recovery was low. Data reported per ADEQ policy 0154.000.
Method/analyte
discrepancies:
T1 =
Method promulgated by EPA, but not by ADHS at this time.
T2 =
Cited ADHS licensed method does not contain this analyte as part of method
compound list.
T3 =
Method not promulgated either by EPA or ADHS.
T4 =
Tentatively identified compound. Concentration is estimated and based on the
closest internal standard.
Calibration
verification:
V1 =
CCV recovery was above method acceptance limits. This target analyte was not
detected in the sample.
V2 =
CCV recovery was above method acceptance limits. This target analyte was
detected in the sample. The sample could not be reanalyzed due to insufficient
sample.
V3 =
CCV recovery was above method acceptance limits. This target analyte was
detected in the sample, but the sample was not reanalyzed. See case narrative.
V4 =
CCV recovery was below method acceptance limits. The sample could not be
reanalyzed due to insufficient sample.
V5 =
CCV recovery after a group of samples was above acceptance limits. This target
analyte was not detected in the sample. Acceptable per EPA Method 8000B.
V6=
Data reported from one-point calibration criteria per ADEQ policy 0155.000.
V7=
Calibration verification recovery was above the method control limit for this
analyte, however the average % difference or % drift for all the analytes met
method criteria.
V8=
Calibration verification recovery was below the method control limit for this
analyte, however the average % difference or % drift for all the analytes met
method criteria.
Calibration:
W1=
The % RSD for this compound was above 15%. The average % RSD for all
compounds in the calibration met the 15% criteria as specified in EPA method
8000B. |